Real-time functionalization and biosensing in subwavelength grating bimodal waveguides

Handle

https://riunet.upv.es/handle/10251/179823

Cita bibliográfica

Torrijos-Morán, L.; Martinez-Perez, P.; García-Rupérez, J. (2020). Real-time functionalization and biosensing in subwavelength grating bimodal waveguides. 1-3. https://riunet.upv.es/handle/10251/179823

Titulación

Resumen

[EN] In this paper, we present for the first time experimental results of biosensing processes by using subwavelength grating bimodal waveguides as a refractive index sensors. We demonstrate the detection of 10ppm concentration of Bovine Serum Albumin (BSA) with a 190pm wavelength shift after a functionalization of the sensor surface with a layer of protein A/G in a continuous flowing system. Real-time data of each step is obtained, validating these kind of novel and highly sensitive SWG sensors for biosensing purposes.

Palabras clave

Subwavelength grating structures, Bimodal waveguides, Refractive-index biosensors, BSA detection, Integrated photonic devices

ISSN

ISBN

Fuente

Proceedings of the 22nd European Conference on Integrated Optics (ECIO 2020)

DOI

Versión del editor

https://www.ecio-conference.org/

dc.description.uri