Mostrar el registro sencillo del ítem
dc.contributor.author | Ródenas Diago, José | es_ES |
dc.contributor.author | Gallardo Bermell, Sergio | es_ES |
dc.contributor.author | Ballester Pérez, Silvia | es_ES |
dc.contributor.author | Primault, Virginie | es_ES |
dc.contributor.author | Ortiz Moragón, Josefina | es_ES |
dc.date.accessioned | 2018-06-07T04:26:10Z | |
dc.date.available | 2018-06-07T04:26:10Z | |
dc.date.issued | 2007 | es_ES |
dc.identifier.issn | 0168-583X | es_ES |
dc.identifier.uri | http://hdl.handle.net/10251/103496 | |
dc.description.abstract | [EN] A gamma spectrometer including an HP Ge detector is commonly used for environmental radioactivity measurements. The efficiency of the detector should be calibrated for each geometry considered. Simulation of the calibration procedure with a validated computer program is an important auxiliary tool for environmental radioactivity laboratories. The MCNP code based on the Monte Carlo method has been applied to simulate the detection process in order to obtain spectrum peaks and determine the efficiency curve for each modelled geometry. The source used for measurements was a calibration mixed radionuclide gamma reference solution, covering a wide energy range (50 2000 keV). Two measurement geometries Marinelli beaker and Petri boxes as well as different materials water, charcoal, sand containing the source have been considered. Results obtained from the Monte Carlo model have been compared with experimental measurements in the laboratory in order to validate the model. | es_ES |
dc.language | Inglés | es_ES |
dc.publisher | Elsevier | es_ES |
dc.relation.ispartof | Nuclear Instruments and Methods in Physics Research Section B Beam Interactions with Materials and Atoms | es_ES |
dc.rights | Reserva de todos los derechos | es_ES |
dc.subject | Detector calibration | es_ES |
dc.subject | Monte Carlo simulation | es_ES |
dc.subject | HP Ge detector | es_ES |
dc.subject | Germanium dead layer | es_ES |
dc.subject | Environmental radioactivity | es_ES |
dc.subject | Gamma spectrometry | es_ES |
dc.subject.classification | INGENIERIA NUCLEAR | es_ES |
dc.title | Application of the Monte Carlo method to the analysis of measurement geometries for the calibration of a HP Ge detector in an environmental radioactivity laboratory | es_ES |
dc.type | Artículo | es_ES |
dc.type | Comunicación en congreso | es_ES |
dc.identifier.doi | 10.1016/j.nimb.2007.04.210 | es_ES |
dc.rights.accessRights | Cerrado | es_ES |
dc.contributor.affiliation | Universitat Politècnica de València. Departamento de Ingeniería Química y Nuclear - Departament d'Enginyeria Química i Nuclear | es_ES |
dc.description.bibliographicCitation | Ródenas Diago, J.; Gallardo Bermell, S.; Ballester Pérez, S.; Primault, V.; Ortiz Moragón, J. (2007). Application of the Monte Carlo method to the analysis of measurement geometries for the calibration of a HP Ge detector in an environmental radioactivity laboratory. Nuclear Instruments and Methods in Physics Research Section B Beam Interactions with Materials and Atoms. 263(1):144-148. doi:10.1016/j.nimb.2007.04.210 | es_ES |
dc.description.accrualMethod | S | es_ES |
dc.relation.conferencename | 6th International Topical Meeting on Industrial Radiation and Radioisotope Measurement Applications | es_ES |
dc.relation.conferencedate | June 20-24,2005 | es_ES |
dc.relation.conferenceplace | Hamilton, Ontario, Canadá | es_ES |
dc.relation.publisherversion | http://doi.org/10.1016/j.nimb.2007.04.210 | es_ES |
dc.description.upvformatpinicio | 144 | es_ES |
dc.description.upvformatpfin | 148 | es_ES |
dc.type.version | info:eu-repo/semantics/publishedVersion | es_ES |
dc.description.volume | 263 | es_ES |
dc.description.issue | 1 | es_ES |
dc.relation.pasarela | S\231760 | es_ES |