Enhancing automatic closed-loop glucose control in type 1 diabetes with an adaptive meal bolus calculator - in silico evaluation under intra- day variability

RiuNet: Repositorio Institucional de la Universidad Politécnica de Valencia

Compartir/Enviar a

Citas

Estadísticas

  • Estadisticas de Uso

Enhancing automatic closed-loop glucose control in type 1 diabetes with an adaptive meal bolus calculator - in silico evaluation under intra- day variability

Mostrar el registro sencillo del ítem

Ficheros en el ítem

dc.contributor.author Herrero, P. es_ES
dc.contributor.author Bondía Company, Jorge es_ES
dc.contributor.author Adewuji, O. es_ES
dc.contributor.author Pesl, P. es_ES
dc.contributor.author El-Sharkawy, M. es_ES
dc.contributor.author Reddy, M. es_ES
dc.contributor.author Toumazou, C. es_ES
dc.contributor.author Oliver, N. es_ES
dc.contributor.author Georgiou, P. es_ES
dc.date.accessioned 2019-12-19T21:02:23Z
dc.date.available 2019-12-19T21:02:23Z
dc.date.issued 2017 es_ES
dc.identifier.issn 0169-2607 es_ES
dc.identifier.uri http://hdl.handle.net/10251/133378
dc.description.abstract [EN] Background and Objective: Current prototypes of closed-loop systems for glucose control in type 1 diabetes mellitus, also referred to as artificial pancreas systems, require a pre-meal insulin bolus to compensate for delays in subcutaneous insulin absorption in order to avoid initial post-prandial hyperglycemia. Computing such a meal bolus is a challenging task due to the high intra-subject variability of insulin requirements. Most closed-loop systems compute this pre-meal insulin dose by a standard bolus calculation, as is commonly found in insulin pumps. However, the performance of these calculators is limited due to a lack of adaptiveness in front of dynamic changes in insulin requirements. Despite some initial attempts to include adaptation within these calculators, challenges remain. Methods: In this paper we present a new technique to automatically adapt the meal-priming bolus within an artificial pancreas. The technique consists of using a novel adaptive bolus calculator based on Case-Based Reasoning and Run-To-Run control, within a closed-loop controller. Coordination between the adaptive bolus calculator and the controller was required to achieve the desired performance. For testing purposes, the clinically validated Imperial College Artificial Pancreas controller was employed. The proposed system was evaluated against itself but without bolus adaptation. The UVa-Padova T1DM v3.2 system was used to carry out a three-month in silico study on 11 adult and 11 adolescent virtual subjects taking into account inter-and intra-subject variability of insulin requirements and uncertainty on carbohydrate intake. Results: Overall, the closed-loop controller enhanced by an adaptive bolus calculator improves glycemic control when compared to its non-adaptive counterpart. In particular, the following statistically significant improvements were found (non-adaptive vs. adaptive). Adults: mean glucose 142.2 ± 9.4 vs. 131.8 ± 4.2 mg/dl; percentage time in target [70, 180] mg/dl, 82.0 ± 7.0 vs. 89.5 ± 4.2; percentage time above target 17.7 ± 7.0 vs. 10.2 ± 4.1. Adolescents: mean glucose 158.2 ± 21.4 vs. 140.5 ± 13.0 mg/dl; percentage time in target, 65.9 ± 12.9 vs. 77.5 ± 12.2; percentage time above target, 31.7 ± 13.1 vs. 19.8 ± 10.2. Note that no increase in percentage time in hypoglycemia was observed. es_ES
dc.description.sponsorship This project has been funded by the Welcome Trust. es_ES
dc.language Inglés es_ES
dc.publisher Elsevier es_ES
dc.relation.ispartof Computer Methods and Programs in Biomedicine es_ES
dc.rights Reserva de todos los derechos es_ES
dc.subject Artificial pancreas es_ES
dc.subject Diabetes es_ES
dc.subject Case-based reasoning es_ES
dc.subject Run-to-Run control es_ES
dc.subject.classification INGENIERIA DE SISTEMAS Y AUTOMATICA es_ES
dc.title Enhancing automatic closed-loop glucose control in type 1 diabetes with an adaptive meal bolus calculator - in silico evaluation under intra- day variability es_ES
dc.type Artículo es_ES
dc.identifier.doi 10.1016/j.cmpb.2017.05.010 es_ES
dc.rights.accessRights Abierto es_ES
dc.contributor.affiliation Universitat Politècnica de València. Departamento de Ingeniería de Sistemas y Automática - Departament d'Enginyeria de Sistemes i Automàtica es_ES
dc.description.bibliographicCitation Herrero, P.; Bondía Company, J.; Adewuji, O.; Pesl, P.; El-Sharkawy, M.; Reddy, M.; Toumazou, C.... (2017). Enhancing automatic closed-loop glucose control in type 1 diabetes with an adaptive meal bolus calculator - in silico evaluation under intra- day variability. Computer Methods and Programs in Biomedicine. 146:125-131. https://doi.org/10.1016/j.cmpb.2017.05.010 es_ES
dc.description.accrualMethod S es_ES
dc.relation.publisherversion https://doi.org/10.1016/j.cmpb.2017.05.010 es_ES
dc.description.upvformatpinicio 125 es_ES
dc.description.upvformatpfin 131 es_ES
dc.type.version info:eu-repo/semantics/publishedVersion es_ES
dc.description.volume 146 es_ES
dc.relation.pasarela S\353469 es_ES
dc.contributor.funder Wellcome Trust


Este ítem aparece en la(s) siguiente(s) colección(ones)