Martínez-Mateu, L.; Romero Pérez, L.; Saiz Rodríguez, FJ.; Berenfeld, O. (2019). Far-field contributions in multi-electrodes atrial recordings blur distinction between anatomical and functional reentries and may cause imaginary phase singularities A computational study. Computers in Biology and Medicine. 108:276-287. https://doi.org/10.1016/j.compbiomed.2019.02.022
Por favor, use este identificador para citar o enlazar este ítem: http://hdl.handle.net/10251/141440
Título:
|
Far-field contributions in multi-electrodes atrial recordings blur distinction between anatomical and functional reentries and may cause imaginary phase singularities A computational study
|
Autor:
|
Martínez-Mateu, Laura
Romero Pérez, Lucia
Saiz Rodríguez, Francisco Javier
Berenfeld, Omer
|
Entidad UPV:
|
Universitat Politècnica de València. Departamento de Ingeniería Electrónica - Departament d'Enginyeria Electrònica
|
Fecha difusión:
|
|
Resumen:
|
[EN] Background
Atrial fibrillation (AF) is the most common cardiac arrhythmia and the most important cause of embolic stroke, requiring new technologies for its better understanding and therapies. Recent approaches to ...[+]
[EN] Background
Atrial fibrillation (AF) is the most common cardiac arrhythmia and the most important cause of embolic stroke, requiring new technologies for its better understanding and therapies. Recent approaches to map the electrical activity during AF with multi-electrode systems aim at localizing patient-specific ablation targets of reentrant patterns. However, there is a critical need to determine the accuracy of those mapping systems. We performed computer simulations as a numerical approach of systematically evaluating the influence of far-field sources on the electrical recordings and detection of rotors.
Methods
We constructed 2 computer models of atrial tissue: (i) a 2D sheet model with varying non-active cells area in its center, and (ii) a whole realistic 3D atrial model. Phase maps were built based on the Hilbert transform of the unipolar electrograms recorded by virtual 2D and 3D multi-electrode systems and rotors were tracked through phase singularities detections.
Results
Analysis of electrograms recorded away from the 2D atrial model shows that the larger the distance between an electrode and the tissue model, the stronger the far-field sources contribution to the electrogram is. Importantly, even if an electrode is positioned in contact with the tissue, the electrogram contains significant contributions from distal sources that blur the distinction between anatomical and functional reentries. Moreover, when mapping the 3D atrial model, remote activity generated false phase singularities at locations without local reentrant excitation patterns.
Conclusions
Far-field contributions to electrograms during AF reduce the accuracy of detecting and interpreting reentrant activity.
[-]
|
Palabras clave:
|
Mapping
,
Electrogram
,
Phase analysis
,
Far-field
,
Rotors
,
Reentry
|
Derechos de uso:
|
Reconocimiento - No comercial - Sin obra derivada (by-nc-nd)
|
Fuente:
|
Computers in Biology and Medicine. (issn:
0010-4825
)
|
DOI:
|
10.1016/j.compbiomed.2019.02.022
|
Editorial:
|
Elsevier
|
Versión del editor:
|
https://doi.org/10.1016/j.compbiomed.2019.02.022
|
Código del Proyecto:
|
info:eu-repo/grantAgreement/NIH//R01HL118304/
info:eu-repo/grantAgreement/GVA//PROMETEO%2F2016%2F088/ES/MODELOS COMPUTACIONALES PERSONALIZADOS MULTI-ESCALA PARA LA OPTIMIZACION DEL DIAGNOSTICO Y TRATAMIENTO DE ARRITMIAS CARDIACAS (PERSONALISED DIGITAL HEART)/
info:eu-repo/grantAgreement/MINECO//DPI2016-75799-R/ES/TECNOLOGIAS COMPUTACIONALES PARA LA OPTIMIZACION DE TERAPIAS PERSONALIZADAS DE PATOLOGIAS AURICULARES Y VENTRICULARES/
|
Agradecimientos:
|
This work was supported in part by Programa Prometeu de la Conselleria d'Educacio, Formacio I Ocupacio de la Generalitat Valenciana, award number PROMETEU/2016/088; Plan Estatal de Investigacion Cientifica y Tecnica y de ...[+]
This work was supported in part by Programa Prometeu de la Conselleria d'Educacio, Formacio I Ocupacio de la Generalitat Valenciana, award number PROMETEU/2016/088; Plan Estatal de Investigacion Cientifica y Tecnica y de Innovacion 2013-2016 del Ministerio de Economia, Industria y Competitividad of Spain, Agencia Estatal de Investigacion and the European Commission (European Regional Development Funds - ERDF -FEDER), award number DPI2016-75799-R; The National Heart, Lung, and Blood Institute grant ROl-HL118304; the Gelman Award from the Cardiovascular Division at the University of Michigan; and the Coulter Program Award from the Dept. of Biomed Eng. at the University of Michigan.
[-]
|
Tipo:
|
Artículo
|