Bellido, P.; Seimetz, M.; Soriano, A.; Huertas, C.; García Lopez, J.; Jimenez-Ramos, MC.; Fernandez, B.... (2013). Dosimetric Calibration of Radiochromic Film For Laser-accelerated Proton Beams. IEEE. 20-23. https://doi.org/10.1109/NSSMIC.2013.6829806
Por favor, use este identificador para citar o enlazar este ítem: http://hdl.handle.net/10251/167117
Título:
|
Dosimetric Calibration of Radiochromic Film For Laser-accelerated Proton Beams
|
Autor:
|
Bellido, P.
Seimetz, Michael
Soriano, A.
Huertas, C.
García Lopez, J.
Jimenez-Ramos, M. C:
Fernandez, B.
Conde, P.
Crespo, E.
González Martínez, Antonio Javier
Hernández, L.
Iborra, A.
Martos, F.
Moliner, L.
Rigla, J. P.
Rodríguez-Álvarez, M. J.
Sánchez, F.
Vidal San Sebastian, Luis Fernando
Benlloch Baviera, Jose María
|
Entidad UPV:
|
Universitat Politècnica de València. Departamento de Matemática Aplicada - Departament de Matemàtica Aplicada
Universitat Politècnica de València. Instituto de Instrumentación para Imagen Molecular - Institut d'Instrumentació per a Imatge Molecular
Universitat Politècnica de València. Instituto Universitario Mixto de Biología Molecular y Celular de Plantas - Institut Universitari Mixt de Biologia Molecular i Cel·lular de Plantes
|
Fecha difusión:
|
|
Resumen:
|
[Otros] When an ultra-intense and ultra-short laser pulse
interacts with solid matter a fraction of the laser pulse can be
converted into kinetic energy of a beam of charged particles.
Radiochromic film (RCF), widely ...[+]
[Otros] When an ultra-intense and ultra-short laser pulse
interacts with solid matter a fraction of the laser pulse can be
converted into kinetic energy of a beam of charged particles.
Radiochromic film (RCF), widely used as radiation detector in
the field of conventional radiotherapy, can be used as detector
for laser-accelerated protons. If used in stack configuration it is
a useful and versatile tool to obtain 2D spatial distribution and
energetic information of proton beams.
In order to obtain dosimetric information from RCF it must
be properly calibrated. Irradiating film pieces under well known
conditions allows us to establish a relation between the optical
density (OD) of the radiochromic film, which is measured
through a flat bed scanner operating in transmission mode, and
the deposited energy in the active layer. A calibration curve
over a large dynamic range (3 orders of magnitude) has been
obtained for few MeV protons. Our calibration process has been
performed at the Spanish National Accelerator Center at Sevilla.
We have irradiated several areas of a single RCF with a constant
50 pA beam current and fixed 4 MeV energy from a 3 MV
tandem accelerator. We have calculated the deposited energy in
the films under the same conditions. We demonstrate that this
technique can be used to measure the spectrum and total energy
of a laser-accelerated mixed-energy proton beam.
This detector has been calibrated for a near future application
at the Center of Pulsed, Ultra-short, Ultra-intense Lasers (CLPU)
at Salamanca (Spain).
We present the calibration procedure and results, the design
optimization, and a comparison with similar experiments.
[-]
|
Derechos de uso:
|
Reserva de todos los derechos
|
ISBN:
|
978-1-4799-0534-8
|
Fuente:
|
2013 IEEE Nuclear Science Symposium and Medical Imaging Conference (2013 NSS/MIC).
|
DOI:
|
10.1109/NSSMIC.2013.6829806
|
Editorial:
|
IEEE
|
Versión del editor:
|
https://doi.org/10.1109/NSSMIC.2013.6829806
|
Título del congreso:
|
IEEE Nuclear Science Symposium and Medical Imaging Conference, and Room-Temperature Semiconductor X-Ray and Gamma-Ray Detectors Workshop (2013 IEEE NSS/MIC/RTSD)
|
Lugar del congreso:
|
Seoul, Korea
|
Fecha congreso:
|
Octubre 27-Noviembre 02,2013
|
Código del Proyecto:
|
info:eu-repo/grantAgreement/GVA//PROMETEOII%2F2013%2F010/ES/Desarrollo de nuevas terapias para tumores cerebrales/
info:eu-repo/grantAgreement/GVA//ISIC2011%2F013/
info:eu-repo/grantAgreement/MICINN//FIS2010-21216-C02-01/ES/DESARROLLO DEL DETECTOR PET%2FRM PARA DIAGNOSTICO DE ENFERMEDADES NEURODEGENERATIVAS./
|
Agradecimientos:
|
Project funded by the Spanish Ministry of Economy and Competitiveness and co-funded with FEDERs funds within the INNPACTO 2011 program. This work was supported by the Spanish Plan Nacional de Investigacion Cientifica, ...[+]
Project funded by the Spanish Ministry of Economy and Competitiveness and co-funded with FEDERs funds within the INNPACTO 2011 program. This work was supported by the Spanish Plan Nacional de Investigacion Cientifica, Desarrollo e InnovacionTecnologica (I+D+i) under Grant No. FIS2010-21216-CO2-01 and the Valencian Local Government under Grants PROMETEOII/2013/010 and ISIC 2011/013.
[-]
|
Tipo:
|
Comunicación en congreso
Capítulo de libro
|