Effect of intracardiac blood flow pulsatility during radiofrequency cardiac ablation: computer modeling study

RiuNet: Repositorio Institucional de la Universidad Politécnica de Valencia

Compartir/Enviar a

Citas

Estadísticas

  • Estadisticas de Uso

Effect of intracardiac blood flow pulsatility during radiofrequency cardiac ablation: computer modeling study

Mostrar el registro sencillo del ítem

Ficheros en el ítem

dc.contributor.author Parés, Cristina es_ES
dc.contributor.author Berjano, Enrique es_ES
dc.contributor.author González-Suárez, Ana es_ES
dc.date.accessioned 2021-11-05T14:06:39Z
dc.date.available 2021-11-05T14:06:39Z
dc.date.issued 2021-01-01 es_ES
dc.identifier.issn 0265-6736 es_ES
dc.identifier.uri http://hdl.handle.net/10251/176249
dc.description.abstract [EN] Purpose To assess the effect of intracardiac blood flow pulsatility on tissue and blood distributions during radiofrequency (RF) cardiac ablation (RFCA). Methods A three-dimensional computer model was used to simulate constant power ablations with an irrigated-tip electrode and three possible catheter orientations (perpendicular, parallel and 45 degrees). Continuous flow and three different pulsatile flow profiles were considered, with four average blood velocity values: 3, 5.5, 8.5 and 24.4 cm/s. The 50 degrees C contour was used to assess thermal lesion size. Results The differences in lesion size between continuous flow and the different pulsatile flow profiles were always less than 1 mm. As regards maximum tissue temperature, the differences between continuous and pulsatile flow were always less than 1 degrees C, with slightly higher differences in maximum blood temperature, but never over 6 degrees C. While the progress of maximum tissue temperature was identical for continuous and pulsatile flow, maximum blood temperature with the pulsatile profile showed small amplitude oscillations associated with blood flow pulsatility. Conclusions The findings show that intracardiac blood pulsatility has a negligible effect on lesion size and a very limited impact on maximum tissue and blood temperatures, which suggests that future experimental studies based on ex vivo or in silico models can ignore pulsatility in intracardiac blood flow. es_ES
dc.description.sponsorship This work was supported by the Spanish Ministerio de Ciencia, Innovacion y Universidades under 'Programa Estatal de IthornDthorni Orientada a los Retos de la Sociedad', Grant No. 'RTI2018-094357-B-C21'. es_ES
dc.language Inglés es_ES
dc.publisher Taylor & Francis es_ES
dc.relation.ispartof International Journal of Hyperthermia es_ES
dc.rights Reconocimiento (by) es_ES
dc.subject Blood flow es_ES
dc.subject Cardiac ablation es_ES
dc.subject Computer model es_ES
dc.subject Pulsatility es_ES
dc.subject Radiofrequency ablation es_ES
dc.subject.classification TECNOLOGIA ELECTRONICA es_ES
dc.title Effect of intracardiac blood flow pulsatility during radiofrequency cardiac ablation: computer modeling study es_ES
dc.type Artículo es_ES
dc.identifier.doi 10.1080/02656736.2021.1890240 es_ES
dc.relation.projectID info:eu-repo/grantAgreement/AEI/Plan Estatal de Investigación Científica y Técnica y de Innovación 2017-2020/RTI2018-094357-B-C21/ES/MODELADO Y EXPERIMENTACION PARA TERAPIAS ABLATIVAS INNOVADORAS/ es_ES
dc.rights.accessRights Abierto es_ES
dc.contributor.affiliation Universitat Politècnica de València. Departamento de Ingeniería Electrónica - Departament d'Enginyeria Electrònica es_ES
dc.description.bibliographicCitation Parés, C.; Berjano, E.; González-Suárez, A. (2021). Effect of intracardiac blood flow pulsatility during radiofrequency cardiac ablation: computer modeling study. International Journal of Hyperthermia. 38(1):316-325. https://doi.org/10.1080/02656736.2021.1890240 es_ES
dc.description.accrualMethod S es_ES
dc.relation.publisherversion https://doi.org/10.1080/02656736.2021.1890240 es_ES
dc.description.upvformatpinicio 316 es_ES
dc.description.upvformatpfin 325 es_ES
dc.type.version info:eu-repo/semantics/publishedVersion es_ES
dc.description.volume 38 es_ES
dc.description.issue 1 es_ES
dc.identifier.pmid 33627008 es_ES
dc.relation.pasarela S\429264 es_ES
dc.contributor.funder AGENCIA ESTATAL DE INVESTIGACION es_ES
upv.costeAPC 1540 es_ES


Este ítem aparece en la(s) siguiente(s) colección(ones)

Mostrar el registro sencillo del ítem