Mostrar el registro sencillo del ítem
dc.contributor.author | Soriano Asensi, Antonio | es_ES |
dc.contributor.author | González Martínez, Antonio Javier | es_ES |
dc.contributor.author | Sánchez Martínez, Filomeno | es_ES |
dc.contributor.author | Conde, Pablo | es_ES |
dc.contributor.author | Moliner Martínez, Laura | es_ES |
dc.contributor.author | Orero Palomares, Abel | es_ES |
dc.contributor.author | Rodríguez Álvarez, María José | es_ES |
dc.contributor.author | Vidal, Luis F. | es_ES |
dc.contributor.author | Benlloch Baviera, Jose María | es_ES |
dc.date.accessioned | 2014-07-08T11:05:37Z | |
dc.date.issued | 2013-04 | |
dc.identifier.issn | 0018-9499 | |
dc.identifier.uri | http://hdl.handle.net/10251/38662 | |
dc.description.abstract | [EN] The increase of the detector thickness and the incidence angle of impinging photons permits an enhancement of sensitivity in positron emission tomography (PET) scanners. But also increases the parallax error and leads to a worsening of spatial resolution. Instead of introducing hardware modifications in the readout electronics or in the detector, we propose in this work to model the photon penetration depth in the detector material and to account for this effect during the image reconstruction. The validation of the model was based on experimental measurements with the MAMMI breast dedicated PET. It consists of twelve detector modules of monolithic LYSO scintillators. A point-like source was acquired at several radial positions across the field of view. The performance of the model was analyzed in terms of position accuracy and spatial resolution. Full width at half maximum (FWHM) average improvement values of 1.0 mm (radial), 0.4 mm (tangential), and 0.3 mm (axial) have been measured when the photon penetration depth was taken into account. The use of the model proposed in this work allows us to design PET detectors with improved sensitivity while maintaining the spatial resolution of the scanner. | es_ES |
dc.description.sponsorship | This work was supported by the Spanish Plan Nacional de Investigación Científica, Desarrollo e Innovación Tecnológica (I+D+I) under Grant FIS2010-21216-CO2-01 and Valencian Local Government under Grants PROMETEO 2008/114 and APOSTD 2010/012. | |
dc.format.extent | 7 | es_ES |
dc.language | Inglés | es_ES |
dc.publisher | Institute of Electrical and Electronics Engineers (IEEE) | es_ES |
dc.relation.ispartof | IEEE Transactions on Nuclear Science | es_ES |
dc.rights | Reserva de todos los derechos | es_ES |
dc.subject | Depth of interaction | es_ES |
dc.subject | PET reconstruction | es_ES |
dc.subject | Scintillation detectors | es_ES |
dc.subject.classification | MATEMATICA APLICADA | es_ES |
dc.title | Minimization of parallax error in dedicated breast PET | es_ES |
dc.type | Artículo | es_ES |
dc.type | Comunicación en congreso | |
dc.embargo.lift | 10000-01-01 | |
dc.embargo.terms | forever | es_ES |
dc.identifier.doi | 10.1109/TNS.2012.2226749 | |
dc.relation.projectID | info:eu-repo/grantAgreement/MICINN//FIS2010-21216-C02-01/ES/DESARROLLO DEL DETECTOR PET%2FRM PARA DIAGNOSTICO DE ENFERMEDADES NEURODEGENERATIVAS./ | es_ES |
dc.relation.projectID | info:eu-repo/grantAgreement/GVA//PROMETEO08%2F2008%2F114/ES/Desarrollo de tecnología PET%2FRM para estudio cerebro en humanos/ | es_ES |
dc.relation.projectID | info:eu-repo/grantAgreement/GVA//APOSTD%2F2010%2F012/ | |
dc.rights.accessRights | Cerrado | es_ES |
dc.contributor.affiliation | Universitat Politècnica de València. Instituto de Instrumentación para Imagen Molecular - Institut d'Instrumentació per a Imatge Molecular | es_ES |
dc.contributor.affiliation | Universitat Politècnica de València. Departamento de Matemática Aplicada - Departament de Matemàtica Aplicada | es_ES |
dc.description.bibliographicCitation | Soriano Asensi, A.; González Martínez, AJ.; Sánchez Martínez, F.; Conde, P.; Moliner Martínez, L.; Orero Palomares, A.; Rodríguez Álvarez, MJ.... (2013). Minimization of parallax error in dedicated breast PET. IEEE Transactions on Nuclear Science. 60(2):739-745. https://doi.org/10.1109/TNS.2012.2226749 | es_ES |
dc.description.accrualMethod | S | es_ES |
dc.relation.conferencename | Symposium on Radiation Measurements and Applications (SORMA) | |
dc.relation.conferencedate | May 14-17, 2012 | |
dc.relation.conferenceplace | Oakland, USA | |
dc.relation.publisherversion | http://ieeexplore.ieee.org/xpl/articleDetails.jsp?arnumber=6407482 | es_ES |
dc.description.upvformatpinicio | 739 | es_ES |
dc.description.upvformatpfin | 745 | es_ES |
dc.type.version | info:eu-repo/semantics/publishedVersion | es_ES |
dc.description.volume | 60 | es_ES |
dc.description.issue | 2 | es_ES |
dc.relation.senia | 252960 | |
dc.contributor.funder | Generalitat Valenciana | |
dc.contributor.funder | Ministerio de Ciencia e Innovación | es_ES |