In vivo electrochemical corrosion study of a CoCrMo biomedical alloy in human synovial fluids

Mostrar el registro sencillo del ítem

Ficheros en el ítem

dc.contributor.author Igual Muñoz, Anna Neus es_ES
dc.contributor.author Schwiesau, J. es_ES
dc.contributor.author Jolles, B. M. es_ES
dc.contributor.author Mischler, Stefano es_ES
dc.date.accessioned 2016-10-17T08:56:45Z
dc.date.available 2016-10-17T08:56:45Z
dc.date.issued 2015-07-15
dc.identifier.issn 1742-7061
dc.identifier.uri http://hdl.handle.net/10251/71895
dc.description.abstract The present study was initiated with the aim to assess the in vivo electrochemical corrosion behaviour of CoCrMo biomedical alloys in human synovial fluids in an attempt to identify possible patient or pathology specific effects. For this, electrochemical measurements (open circuit potential OCP, polarization resistance R-p, potentiodynamic polarization curves, electrochemical impedance spectroscopy EIS) were carried out on fluids extracted from patients with different articular pathologies and prosthesis revisions. Those electrochemical measurements could be carried out with outstanding precision and signal stability. The results show that the corrosion behaviour of CoCrMo alloy in synovial fluids not only depends on material reactivity but also on the specific reactions of synovial fluid components, most likely involving reactive oxygen species. In some patients the latter were found to determine the whole cathodic and anodic electrochemical response. Depending on patients, corrosion rates varied significantly between 50 and 750 mg dm(-2) year(-1). es_ES
dc.language Inglés es_ES
dc.publisher Elsevier es_ES
dc.relation.ispartof Acta Biomaterialia es_ES
dc.rights Reserva de todos los derechos es_ES
dc.subject Human synovia es_ES
dc.subject Corrosion es_ES
dc.subject Electrochemistry es_ES
dc.subject CoCrMo alloy es_ES
dc.subject Knee implants es_ES
dc.subject.classification INGENIERIA QUIMICA es_ES
dc.title In vivo electrochemical corrosion study of a CoCrMo biomedical alloy in human synovial fluids es_ES
dc.type Artículo es_ES
dc.identifier.doi 10.1016/j.actbio.2015.03.008
dc.rights.accessRights Cerrado es_ES
dc.contributor.affiliation Universitat Politècnica de València. Departamento de Ingeniería Química y Nuclear - Departament d'Enginyeria Química i Nuclear es_ES
dc.description.bibliographicCitation Igual Muñoz, AN.; Schwiesau, J.; Jolles, BM.; Mischler, S. (2015). In vivo electrochemical corrosion study of a CoCrMo biomedical alloy in human synovial fluids. Acta Biomaterialia. 21:228-236. doi:10.1016/j.actbio.2015.03.008 es_ES
dc.description.accrualMethod S es_ES
dc.relation.publisherversion http://dx.doi.org/10.1016/j.actbio.2015.03.008 es_ES
dc.description.upvformatpinicio 228 es_ES
dc.description.upvformatpfin 236 es_ES
dc.type.version info:eu-repo/semantics/publishedVersion es_ES
dc.description.volume 21 es_ES
dc.relation.senia 304174 es_ES
dc.identifier.eissn 1878-7568


Este ítem aparece en la(s) siguiente(s) colección(ones)

Mostrar el registro sencillo del ítem