Mostrar el registro sencillo del ítem
dc.contributor.author | Gerardy, I. | es_ES |
dc.contributor.author | Ródenas Diago, José | es_ES |
dc.contributor.author | van Dycke, M. | es_ES |
dc.contributor.author | Gallardo Bermell, Sergio | es_ES |
dc.contributor.author | Ceccolini, E. | es_ES |
dc.date.accessioned | 2018-01-17T10:21:56Z | |
dc.date.available | 2018-01-17T10:21:56Z | |
dc.date.issued | 2010 | es_ES |
dc.identifier.issn | 0969-8043 | es_ES |
dc.identifier.uri | http://hdl.handle.net/10251/94988 | |
dc.description.abstract | [EN] A gynaecological applicator consisting of a metallic intra-uterine tube with a plastic vaginal applicator and an HDR Ir-192 source have been simulated with MCNP5 (Monte Carlo code). A solid phantom has been designed to perform measurements around the applicator with radiochromic films. The isodose curves obtained are compared with curves calculated with the F4MESH tally of MCNP5 with a good agreement. A pinpoint ionization chamber has been used to evaluate dose at some reference points. (c) 2009 Elsevier Ltd. All rights reserved. | es_ES |
dc.language | Inglés | es_ES |
dc.publisher | Elsevier | es_ES |
dc.relation.ispartof | Applied Radiation and Isotopes | es_ES |
dc.rights | Reserva de todos los derechos | es_ES |
dc.subject | Brachytherapy | es_ES |
dc.subject | MCNP5 | es_ES |
dc.subject | Dosimetry | es_ES |
dc.subject | Radiochromic films | es_ES |
dc.subject | Pinpoint chamber | es_ES |
dc.subject.classification | INGENIERIA NUCLEAR | es_ES |
dc.title | Dosimetric characterization of a brachytherapy applicator using MCNP5 modelisation and in-phantom measurements | es_ES |
dc.type | Artículo | es_ES |
dc.identifier.doi | 10.1016/j.apradiso.2009.10.018 | es_ES |
dc.rights.accessRights | Abierto | es_ES |
dc.contributor.affiliation | Universitat Politècnica de València. Departamento de Ingeniería Química y Nuclear - Departament d'Enginyeria Química i Nuclear | es_ES |
dc.description.bibliographicCitation | Gerardy, I.; Ródenas Diago, J.; Van Dycke, M.; Gallardo Bermell, S.; Ceccolini, E. (2010). Dosimetric characterization of a brachytherapy applicator using MCNP5 modelisation and in-phantom measurements. Applied Radiation and Isotopes. 68(4):735-737. doi:10.1016/j.apradiso.2009.10.018 | es_ES |
dc.description.accrualMethod | S | es_ES |
dc.relation.publisherversion | https://doi.org/10.1016/j.apradiso.2009.10.018 | es_ES |
dc.description.upvformatpinicio | 735 | es_ES |
dc.description.upvformatpfin | 737 | es_ES |
dc.type.version | info:eu-repo/semantics/publishedVersion | es_ES |
dc.description.volume | 68 | es_ES |
dc.description.issue | 4 | es_ES |
dc.relation.pasarela | S\40507 | es_ES |