A non-parametric validation framework for photoplethysmography-based heart rate monitoring: a proof-of-concept study using the two-sample Kolmogorov–Smirnov test

Handle

https://riunet.upv.es/handle/10251/232181

Cita bibliográfica

Cruz-Salgado, J.; Bautista-López, RZ.; Yañez-Mendiola, J.; Ruelas-Santoyo, EA.; Miguel-Andrés, I.; Jiménez-García, JA. (2026). A non-parametric validation framework for photoplethysmography-based heart rate monitoring: a proof-of-concept study using the two-sample Kolmogorov–Smirnov test. Journal of Applied Research in Technology & Engineering. 7(1):58-70. https://riunet.upv.es/handle/10251/232181

Titulación

Resumen

[EN] This article introduces a statistical approach for comparing heart rate measurements obtained from two photoplethysmography (PPG) signals: one recorded with a commercial oximeter and the other acquired using a device based on photoplethysmography with synchronous detection. The study applies the two-sample Kolmogorov Smirnov test as a robust and versatile method for comparing the distributions of PPG signals. By integrating the two-sample Kolmogorov Smirnov test into the validation process of cardiac pulse measurement devices, the work demonstrates its effectiveness in enhancing the accuracy and reliability of biomedical signal analysis. Results show that, when comparing signals against calibrated reference devices and visualizing cumulative distribution functions, the two-sample Kolmogorov Smirnov test is capable of detecting subtle differences in signal behavior. This innovative use of the two-sample Kolmogorov Smirnov test provides valuable insights for the design and validation of biomedical signal processing systems and contributes to the advancement of non-invasive health monitoring technologies.

Fuente

Journal of Applied Research in Technology & Engineering

DOI

Versión del editor

Enlaces relacionados

URL