Quality assurance applied to digital radiographic equipment by developed software for phantom images

RiuNet: Repositorio Institucional de la Universidad Politécnica de Valencia

Compartir/Enviar a

Citas

Estadísticas

  • Estadisticas de Uso

Quality assurance applied to digital radiographic equipment by developed software for phantom images

Mostrar el registro sencillo del ítem

Ficheros en el ítem

dc.contributor.author Mayo Nogueira, Patricia es_ES
dc.contributor.author Ródenas Escribá, Francisco de Asís es_ES
dc.contributor.author Campayo, J. M. es_ES
dc.contributor.author Marin, B. es_ES
dc.contributor.author Verdú Martín, Gumersindo Jesús es_ES
dc.date.accessioned 2013-04-22T12:44:50Z
dc.date.issued 2011
dc.identifier.issn 0029-5450
dc.identifier.uri http://hdl.handle.net/10251/28110
dc.description.abstract [EN] The assessment and control of image quality is a fundamental task associated with good practice to guarantee a suitable diagnosis by the radiologist. The need for image quality assessments in radiography is well established, and the use of test phantoms is a common method for this purpose. In this work we present a developed tool that consists of a specific phantom (named RACON) that is used for acceptance and constancy test in order to analyze the image obtained by digital radiographic equipment, software (named SoftRACON) for automated image analysis with digital processing techniques, and a database to store test phantom images and the scoring results. The main objective is to characterize the constancy of the radiographic imaging chain and guarantee acceptable image quality, related to well-functioning of the radiographic equipment. Therefore, the application presented in this work is sensitive enough to the operating conditions of the radiographic digital equipment and allows the assessment of the imaging system quality and, consequently, increases the objectivity (accuracy) in the evaluation of the image. es_ES
dc.language Inglés es_ES
dc.publisher American Nuclear Society es_ES
dc.relation.ispartof Nuclear Technology es_ES
dc.rights Reconocimiento - No comercial (by-nc) es_ES
dc.subject Digital image es_ES
dc.subject Quality assurance es_ES
dc.subject Radiographic phantom es_ES
dc.subject Automated image analysis es_ES
dc.subject Digital equipment es_ES
dc.subject Good practices es_ES
dc.subject Image quality assessment es_ES
dc.subject Operating condition es_ES
dc.subject Phantom images es_ES
dc.subject Radiographic imaging es_ES
dc.subject Equipment es_ES
dc.subject Image quality es_ES
dc.subject Radiology es_ES
dc.subject Software testing es_ES
dc.subject Quality control es_ES
dc.subject.classification INGENIERIA NUCLEAR es_ES
dc.subject.classification MATEMATICA APLICADA es_ES
dc.title Quality assurance applied to digital radiographic equipment by developed software for phantom images es_ES
dc.type Artículo es_ES
dc.identifier.doi 10.13182/NT11-A12268
dc.rights.accessRights Abierto es_ES
dc.contributor.affiliation Universitat Politècnica de València. Departamento de Matemática Aplicada - Departament de Matemàtica Aplicada es_ES
dc.contributor.affiliation Universitat Politècnica de València. Departamento de Ingeniería Química y Nuclear - Departament d'Enginyeria Química i Nuclear es_ES
dc.description.bibliographicCitation Mayo Nogueira, P.; Ródenas Escribá, FDA.; Campayo, JM.; Marin, B.; Verdú Martín, GJ. (2011). Quality assurance applied to digital radiographic equipment by developed software for phantom images. Nuclear Technology. 175(1):48-52. https://doi.org/10.13182/NT11-A12268 es_ES
dc.description.accrualMethod S es_ES
dc.relation.publisherversion http://dx.doi.org/10.13182/NT11-A12268 es_ES
dc.description.upvformatpinicio 48 es_ES
dc.description.upvformatpfin 52 es_ES
dc.type.version info:eu-repo/semantics/publishedVersion es_ES
dc.description.volume 175 es_ES
dc.description.issue 1 es_ES
dc.relation.senia 216602


Este ítem aparece en la(s) siguiente(s) colección(ones)

Mostrar el registro sencillo del ítem