Calibration of Gamma Ray Impacts in Monolithic-Based Detectors Using Voronoi Diagrams
Fecha
Directores
Editores
Otras autorías
Unidades organizativas
Handle
Cita bibliográfica
Titulación
Resumen
[EN] Molecular imaging systems, such as positron emission tomography (PET), use detectors providing energy and a 3-D interaction position of a gamma ray within a scintillation block. Monolithic crystals are becoming an alternative to crystal arrays in PET. However, calibration processes are required to correct for nonuniformities, mainly produced by the truncation of the scintillation light distribution at the edges. We propose a calibration method based on the Voronoi diagrams. We have used