Calibration of Gamma Ray Impacts in Monolithic-Based Detectors Using Voronoi Diagrams

dc.contributor.affiliationInstituto de Instrumentación para Imagen Molecular
dc.contributor.authorGonzález Martínez, Antonio Javier
dc.contributor.authorFreire-López-Fando, Marta
dc.contributor.authorGonzalez-Montoro, Andreaes_ES
dc.contributor.authorF Sánchez
dc.contributor.authorBenlloch Baviera, Jose María
dc.contributor.funderEuropean Commissiones_ES
dc.contributor.funderMinisterio de Economía y Competitividades_ES
dc.date.accessioned2021-07-08T03:31:09Z
dc.date.available2021-07-08T03:31:09Z
dc.date.issued2020-05es_ES
dc.description.abstract[EN] Molecular imaging systems, such as positron emission tomography (PET), use detectors providing energy and a 3-D interaction position of a gamma ray within a scintillation block. Monolithic crystals are becoming an alternative to crystal arrays in PET. However, calibration processes are required to correct for nonuniformities, mainly produced by the truncation of the scintillation light distribution at the edges. We propose a calibration method based on the Voronoi diagrams. We have used $50 \times 50 \times 15$ mm(3) LYSO blocks coupled to a $12\times 12$ SiPMs array. We have first studied two different interpolation algorithms: 1) weighted average method (WAM) and 2) natural neighbor (NN). We have compared them with an existing calibration based on 1-D monomials. Here, the crystal was laterally black painted and a retroreflector (RR) layer added to the entrance face. The NN exhibited the best results in terms of XY impact position, depth of Interaction, and energy, allowing us to calibrate the whole scintillation volume. Later, the NN interpolation has been tested against different crystal surface treatments, allowing always to correct edge effects. Best energy resolutions were observed when using the reflective layers (12%-14%). However, better linearity was observed with the treatments using black paint. In particular, we obtained the best overall performance when lateral black paint is combined with the RR.en_EN
dc.description.accrualMethodSes_ES
dc.description.bibliographicCitationFreire, M.; Gonzalez-Montoro, A.; Sánchez Martínez, F.; Benlloch Baviera, JM.; González Martínez, AJ. (2020). Calibration of Gamma Ray Impacts in Monolithic-Based Detectors Using Voronoi Diagrams. IEEE Transactions on Radiation and Plasma Medical Sciences. 4(3):350-360. https://doi.org/10.1109/TRPMS.2019.2947716es_ES
dc.description.issue3es_ES
dc.description.sponsorshipThis work was supported in part by the European Research Council through the European Union's Horizon 2020 Research and Innovation Program under Grant 695536, and in part by the Spanish Ministerio de Economia, Industria y Competitividad under Grant TEC2016-79884-C2-1-R.es_ES
dc.description.upvformatpfin360es_ES
dc.description.upvformatpinicio350es_ES
dc.description.volume4es_ES
dc.identifier.doi10.1109/TRPMS.2019.2947716es_ES
dc.identifier.issn2469-7311es_ES
dc.identifier.urihttps://riunet.upv.es/handle/10251/168935
dc.languageIngléses_ES
dc.publisherInstitute of Electrical and Electronics Engineerses_ES
dc.relation.ispartofIEEE Transactions on Radiation and Plasma Medical Scienceses_ES
dc.relation.pasarelaS\432478es_ES
dc.relation.projectIDinfo:eu-repo/grantAgreement/EC/H2020/695536/EU/Innovative PET scanner for dynamic imaging/es_ES
dc.relation.projectIDinfo:eu-repo/grantAgreement/MINECO/Programa Estatal de I+D+i Orientado a los Retos de la Sociedad/TEC2016-79884-C2-1-R/Desarrollo del hardware para sistema de diagnóstico por imagen molecular para corazón en condiciones de estrés/es_ES
dc.relation.publisherversionhttps://doi.org/10.1109/TRPMS.2019.2947716es_ES
dc.rightsReconocimiento (by)es_ES
dc.rights.accessRightsAbiertoes_ES
dc.subjectCalibration processeses_ES
dc.subjectGamma ray detectorses_ES
dc.subjectMonolithic crystalses_ES
dc.subjectPositron emission tomography (PET)es_ES
dc.subjectSiPMes_ES
dc.subjectVoronoi diagramses_ES
dc.titleCalibration of Gamma Ray Impacts in Monolithic-Based Detectors Using Voronoi Diagramses_ES
dc.typeArtículoes_ES
dc.type.versioninfo:eu-repo/semantics/publishedVersiones_ES
dspace.entity.typePublication
person.identifier407345
person.identifier604281
person.identifier468362
person.identifier396595
person.identifier.orcid0000-0001-6073-1436
relation.isAuthorOfPublication3484f0b6-dae3-4eb1-8694-1f189d247de8
relation.isAuthorOfPublication70a3ebf0-c269-4f97-b673-e013307fde6a
relation.isAuthorOfPublication3796e255-4b46-4ab0-b24c-e3f5041efbc4
relation.isAuthorOfPublicationf19a2036-3a9b-49bf-a03b-85480ac291f2
relation.isAuthorOfPublication.latestForDiscovery3796e255-4b46-4ab0-b24c-e3f5041efbc4
relation.isOrgUnitOfPublication2a147664-9b3c-4f73-8ea8-aefbad2ed456
relation.isOrgUnitOfPublication.latestForDiscovery2a147664-9b3c-4f73-8ea8-aefbad2ed456
upv.uuid382ac67f-b91d-4e88-9f74-e2a6c85b1c9bes_ES

Archivos

Bloque original

Mostrando 1 - 1 de 1
Cargando...
Miniatura
Nombre:
Frere;Gonzalez-Montoro;Sánchez - Calbraton of Gamma Ray Impactsn Monolthc-Based Detectors Usng Vo....pdf
Tamaño:
7.18 MB
Formato:
Adobe Portable Document Format
Descripción:
Versión editorial